Download PDFOpen PDF in browser

A Soft Palpation Sensor for Early Detection of Biological Tissue Abnormalities Based on Electrical Impedance Tomography

EasyChair Preprint no. 8171

2 pagesDate: June 1, 2022

Abstract

We propose a soft palpation sensor to detect anomalies in tissue elastic properties, to enable identification of lesions not visible through endoscopic imaging, such as precancerous colorectal tumours. The sensor includes an inflatable unit which distends the tissue analogous to a palpation task, variations on the sensor surface caused by the tissue stiffness is detected by measuring the electrical impedance within the inflatable chamber.  Palpation experiments were performed where the sensor was attached to a robot arm and compared to an industrial force sensor. The sensor was able to detect 4mm diameter nodules 3mm under the surface, which would not have been visible using visual techniques.

Keyphrases: Electrical Impedance Tomography, Palpation Sensor, Soft Robotics

BibTeX entry
BibTeX does not have the right entry for preprints. This is a hack for producing the correct reference:
@Booklet{EasyChair:8171,
  author = {Saina Akhond and James Avery and George Mylonas},
  title = {A Soft Palpation Sensor for Early Detection of Biological Tissue Abnormalities Based on Electrical Impedance Tomography},
  howpublished = {EasyChair Preprint no. 8171},

  year = {EasyChair, 2022}}